FLS C100-300 STAINLESS STEEL CONDUCTIVITY SENSORS


The FLS conductivity sensors with stainless steel electrodes (C100 series) are designed for agriculture application and for light industrial application obviously where sample conditions allow a steel using (water treatment, foods industry and others). This type of sensors are characterized by a considerable ratio performance/price. The combination of temperature sensor with the ATC (Automatic Temperature Compensation) function of the monitor/transmitter allow to get a precise measurement. Moreover a wide number of cell constants grants to choose the best item for specific application. C300 serie has been designed for ultrapure water monitoring (certified cell constant 0,01) and for waste water application (cell constant 10). C300 sensors are completely made in SS ensuring suitability for a wide range of applications.

Description

Applications:

  • Agriculture and fertilizing system
  • Water treatment
  • Foods industry
  • Aquaculture
  • Ultrapure water application: production and use

Main features:

  • Stainless steel measuring surfaces
  • Considerable ratio performance/price
  • Temperature sensor included
  • Wide range of cell constant
  • Rugged sensor body in PP (C100)
  • Sensor in SS completely (C300)
  • C300.001TCCK with certified cell constant

Technical data:

General
• Operating range:

  • C300.001 TC: 0,055 µS to 200 µS (18,2 MΩ to 5 KΩ)
  • C100.01 TC: 0.1 µS to 2000 µS (10 MΩ to 500 Ω)
  • C100.02 TC: 0.2 µS to 4000 µS
  • C100.1 TC: 1 µS to 20000 µS
  • C300.10 TC: 10 µS to 200000 µS

• Temperature compensation device (for TC models):

  • Pt 100 (C100TC), Pt1000 (C300TC)

• Cable length:

  • C100.01, C100.02, C100.1: 5 m bipolaire kabel met

5 mm doorsnede

  • C100.01 TC, C100.02 TC, C100.1 TC: no cable available
  • C300 TC: 3m

• Process connection:

  • C100.01 TC, C100.02 TC, C100.1 TC: ¾” male BSP
  • C300.TC: ¾” male BSP fitting

• Max. werktemperatuur:

  • C100.01, C100.02, C100.1: 80°C (176°F)
  • C100.01 TC, C100.02 TC, C100.1 TC: 80°C (176°F)
  • C300TC: 80°C (PP fitting), 120°C (RVS fitting)

• Max. Working Temperature:

  • C100.01 TC, C100.02 TC, C100.1 TC : 80°C (176°F)
  • C300TC: 80°C (PP fitting), 120°C (SS fitting)

• Max. Working Pressure:

  • C100.01 TC, C100.02 TC, C100.1 TC: 6 bar (85 psi)
  • C300 TC: 7 bar (PP fitting), 13 bar (SS fitting)

• Wetted materials:

  •  body: C100.01 TC, C100.02 TC, C100.1 TC: PP; C300 TC: SS 316
  • measuring surface: stainless steel AISI 316

Standards & Approvals

  • Manufactured under ISO 9001
  • Manufactured under ISO 14001
  • CE • RoHS Compliant
  • GOST R

Specifications

Ask a quotation for this product

Related Products

You may also be interested in the following products:

FLS M9.07 DUAL-PARAMETER CONDUCTIVITY AND FLOW MONITOR & TRANSMITTER

FLS M9.07 DUAL-PARAMETER CONDUCTIVITY AND FLOW MONITOR & TRANSMITTER

FLS M9.05 CONDUCTIVITY MONITOR & TRANSMITTER

FLS M9.05 CONDUCTIVITY MONITOR & TRANSMITTER

© 2012-2019 Ebora Process Automation is a trademark of Ebora BV - 't Veld 3 - S, Heteren, Netherlands - info@ebora.nl